5-year battery life
USB, serial ports
1 second - 12 hours intervals
Rugged aluminum enclosure
Related Links
SITE-LOG LPSE-1

7-channel pulse/state/event data logger

SITE-LOG Logger The SITE-LOG LPSE-1 is 7-channel, battery powered, stand-alone, pulse/state/event data logger. The logger detects electronic or mechanical pulse counts, state changes or events in any of the seven channels and the data are saved in 4MB non-volatile flash memory for later retrieval. It works with gas/water/power meter, rain gauge, flow rate meter and any other digital signal/switch closure output devices.

Powered by configurable channel, internal pull-up and triggering edge selections, the LPSE-1 combines the functionalities of three types of loggers into one compact unit and is suitable for wide industrial applications.

Featuring an aluminum enclosure and conformal coating PCB, the SITE-LOG data logger has excellent performance in the harshest industrial environment. plug & Play USB port and versatile custom equation simplify communications and engineering unit conversion.

Simply plug the logger to computer's USB port, and the software automatically recognizes it and handles the configuration, downloading, graph viewing and more...

Features:
  • 4 megabyte memory size
  • Both USB and serial interfaces with auto baud rate of up to 115200 bps
  • Programmable input channels for pulse, state or event inputs
  • Over 5 years of battery Life
  • Powerful software for configuration, downloading, plotting, analysis and alarm reporting
  • Configurable alarm control/excitation controls
  • Wide sampling interval selections (1 second to 12 hours)
  • Rugged aluminum enclosure
  • Coated PCB to protect against moisture, corrosion and mildew.
Logger Memory Capacity
Sampling Interval Enabled Channel Logging Capacity
1 minute 1 484 days
1 minute 7 69 days
10 seconds 1 80 days
10 seconds 7 11 days
1 second 1 8 days
1 second 7 27 hours

Note:
For State and Event channel, the above capacity is calculated based on a state or event occurs in every sampling period.


Specifications
7-channel pulse/state/event data logger
General
Product Name:
SITE-LOG
Model:
LPSE-1
Inputs
Input Connection:
Pluggable terminal block for seven external channels, excitation controls and alarm outputs
Channels:
Seven external channels record state changes, event or pulse count.
Each channel can be configured as either stage, event or pulse channel based on applications.
For Event and Pulse configuration, each channel's trigger detection can be configured between Normal Open (High to Low) (suggested) and Normal Close (Low to High) (not suggested)
Type of Inputs:
Voltage, solid state switch, coil relay, dry contact, digital TTL signal.
Max. Pulse Count per Interval:
65535 pulse counts
Pulse Accuracy
± 1 pulse
Max. Pulse Rate:
Without external power supply: 10 HZ (20 HZ if only one channel enabled)
With external power supply : 2 KHZ (4 KHZ if only one channel enabled)
Mimimum Pulse Width:
Without external power supply: 35 milliseconds
With external power supply: 0.5 milliseconds
Internal Weak Pull-UP:
100 KOhms. Switch on/off by software
Input Impedance:
>= 1MOhm
Max. Input Voltage:
40 Volts
Trigger High Voltage:
>= 2.7 Volts
Trigger Low Voltage:
<= 0.5 Volts
Alarms:
Channel Alarms:
Two editable alarm thresholds per channel
Alarm Output:
ALARM1 & A2/E terminal strips can be configured as alarm outputs
Alarm-On: MOSFET(N-Channel) switch on
Alarm-Off: MOSFET(N-Channel) switch off
Max Power: 200mA @ 24VDC
With purchase of SiteView software, SITE-LOG can report alarm status to host PC via USB, Modem or Ethernet Device Server.
Alarm-On Delay:
programmable 0 - 10 minutes delay with 1-minute increment.
Alarm Indicator:
On-board LED lights in red when in alarm condition.
On-board Memory:
Size:
4 Mega bytes (2MB readings)
Data Retention:
Over 20 years
Sampling:
Sampling Interval:
1 second to 12 hours user selectable
Logging Mode:
Stop recording or FIFO when memory is full
Logging Activation:
Programmable instant, start delay or field push button activation.
Communications:
Interface:
USB (Mini-USB-B) (USB-A-mini USB-B Cable included)
AUX(RJ11) for direct TTL level communications
With purchase of DeviceServer or MODEM Kit SITE-LOG can be connected to Ethernet or MODEM for remote access.
Baud Rate:
Auto-detect baud rate from 2400 to 115200 bps on both USB and AUX.
Battery:
Power:
Built-in 3.6V Lithium Battery.
Battery Life:
5 years if 1 minute sampling interval or longer, normal open contact
Software:
SiteView: (Sold separately)
Configuration, downloading, plotting, real-time view, custom calibration and custom equation.
Software Requirements:
Computer with 1.0 GHz or faster processor
256 MB Memory or higher
1.0 GB of available hard-drive space or higher
Windows XP with SP2 or later, Vista, Window 7
At least one USB port or one COM port.
Physical:
Material
Aluminum enclosure.
PCB Treatment:
Conformal coating.
Dimension:
88 X 64.2 X 24 mm (3.46 X 2.53 X 0.95 inches).
Weight:
200g
Mounting:
Probe/wall-mouint holes for hanging/mounting.
Others:
LED Indicator
Tri-Color LED: (can be disabled for power saving)
Normal Sampling: green when sampling.
Alarm: red when sampling.
Low Battery: amber when sampling.
Excitation Control:
A2/E terminal strip can be configured as excitation control output for driving the power of connected devices.
Delay Interval settings: 5 to 240 seconds with 5-second increment.
Operating Environment:
-40 ~ +70 °C (-40 °F ~ 158 °F), 0 ~ 95% RH non-condensing.
Clock Accuracy:
± 1 mintue per month.
Approvals:
CE, FCC
SiteView Software Overview Real-Time Strip Chart Graph & Tabular Views Logger Status Window


Model Description Buy