PL-C PRECISE-LOG 8-Channel Current Data Logger
The PRECISE-LOG PL-C is an 8-channel, battery-powered, standalone DC current data logger designed for industrial, laboratory, research, testing, process-monitoring, and long-term data acquisition applications.
The logger provides:
- 8 external DC current input channels
- 16-bit measurement resolution
- USB communication
- 8 MB non-volatile flash memory
- Long-term standalone operation
Each external channel can be independently enabled and configured for the required current input range.
The programmable current ranges are:
- 4–20 mA
- 0–50 mA
The PL-C is the non-WiFi version of the PRECISE-LOG current platform and is well suited to applications requiring standalone recording, direct USB communication, or network access through Smart-Hub.
Multi-Channel Current Measurement
The PL-C provides eight external DC current channels for simultaneous monitoring of industrial transmitters, sensors, process instruments, and other compatible current-output devices.
Each channel can be independently enabled and configured through SiteView.
The available programmable input ranges are:
- 4–20 mA
- 0–50 mA
The logger provides:
- 0.0018% measurement resolution
- ±0.1% FSR accuracy at 25°C
- 12 Ω load resistor
- Over-current protection up to ±100 mA
These characteristics make the PL-C suitable for standard industrial process signals as well as general DC current measurement applications.
High-Resolution Data Logging
The PL-C uses a 16-bit analog-to-digital converter with a measurement resolution of:
0.0018%
This provides high-resolution current measurement for industrial, laboratory, research, validation, and testing applications.
Standalone Operation and Long-Term Recording
The PL-C operates independently after configuration and stores measurements in 8 MB of non-volatile flash memory.
The memory provides storage for approximately 4 million readings and retains recorded data for more than 20 years.
Logging can be configured for:
- Immediate start
- Delayed start
- Field push-button activation
When memory becomes full, the logger can either:
- Stop recording, or
- Continue recording in FIFO mode
The built-in 3.6 V lithium battery provides up to approximately 10 years of operation at a 1-minute sampling interval in standalone mode, depending on operating conditions and configuration.
USB Communication and SiteView Software
The PL-C connects directly to a Windows PC through its Plug & Play USB interface.
SiteView software provides tools for:
- Configuring current channels
- Selecting 4–20 mA or 0–50 mA input ranges
- Configuring logging sessions
- Downloading recorded data
- Viewing real-time current measurements
- Displaying graphs and tables
- Configuring alarm thresholds
- Applying custom calibration
- Applying custom equations
- Converting current signals into engineering units
- Exporting and analyzing recorded data
For local operation, SiteView communicates directly with the PL-C through USB.
For network applications, the PL-C can be connected to a Smart-Hub and accessed remotely by supported Microedge software.
Custom equations are especially useful when a 4–20 mA transmitter represents another engineering quantity such as:
- Pressure
- Flow
- Level
- Temperature
- Humidity
- Position
- Other process variables
Alarm Functions
The PL-C provides configurable alarm functions for current-monitoring applications.
Each channel supports two programmable alarm thresholds.
The onboard tri-color status LED provides visual indication of:
- Normal sampling
- Alarm condition
- Low battery
The LED can be disabled to reduce power consumption.
Remote Access and Network Integration
Although the PL-C is a USB-based data logger, it can also be accessed remotely over an Ethernet network by connecting it to a Microedge Smart-Hub.
Smart-Hub acts as a communication gateway and allows connected USB data loggers to be accessed through:
- SiteView
- SiteView-Web
- SiteView SDK
- Pluto
Smart-Hub also provides:
- RESTful Web API
- Modbus TCP Gateway
This allows PL-C current data to be integrated with third-party software, PLCs, HMIs, SCADA systems, and other industrial monitoring platforms.
Unlike the PL-CW, the PL-C does not contain a built-in WiFi interface and does not operate directly as a Modbus TCP server. Network, RESTful API, and Modbus TCP access for PL-C are provided through Smart-Hub.
Rugged Industrial Design
The PL-C is housed in a durable aluminum enclosure designed for laboratory, industrial, and field applications.
Its operating environment is:
−40°C to +70°C (−40°F to +158°F)
and:
0–95% RH, non-condensing.
KEY FEATURES
8 External DC Current Channels
Provides simultaneous multi-channel current measurement for industrial sensors, transmitters, and process signals.Programmable Current Ranges
Supports 4–20 mA and 0–50 mA input ranges.High Measurement Accuracy
Provides ±0.1% FSR accuracy at 25°C.16-Bit Data Resolution
Provides 0.0018% measurement resolution.12 Ω Input Load
Each current input uses a 12 Ω load resistor.Over-Current Protection
Current inputs are protected up to ±100 mA.8 MB Non-Volatile Flash Memory
Stores approximately 4 million readings for long-term standalone recording.1 Second to 12 Hour Sampling
Wide programmable sampling interval selection for short-term testing and long-term monitoring.Long Battery Life
Up to approximately 10 years at a 1-minute sampling interval in standalone operation.USB Communication
Provides direct local communication with SiteView.SiteView Software Support
Provides logger configuration, data download, real-time monitoring, plotting, calibration, custom equations, alarms, and data analysis.Programmable Alarm Thresholds
Two configurable alarm thresholds are available for each channel.Flexible Logging Modes
Supports immediate start, delayed start, field push-button activation, stop-on-full, and FIFO operation.Remote Network Access through Smart-Hub
Smart-Hub provides Ethernet access for SiteView, SiteView-Web, SiteView SDK, and Pluto.RESTful Web API Integration
Through Smart-Hub, PL-C data can be accessed by compatible third-party software.Modbus TCP Gateway Integration
Smart-Hub can expose connected PL-C data through Modbus TCP for integration with PLCs, HMIs, SCADA systems, and other Modbus TCP masters.Custom Engineering-Unit Conversion
SiteView custom equations can convert current signals into pressure, flow, level, temperature, humidity, and other engineering units.Rugged Aluminum Enclosure
TYPICAL APPLICATIONS
The PL-C is suitable for:
- 4–20 mA transmitter monitoring
- 0–50 mA current measurement
- Industrial process monitoring
- Pressure transmitter monitoring
- Flow transmitter monitoring
- Level transmitter monitoring
- Environmental sensor monitoring
- Industrial sensor monitoring
- Laboratory testing
- Research
- Product validation
- Equipment monitoring
- Long-term unattended current logging
- Remote current monitoring through Smart-Hub
- PLC / SCADA process monitoring through Smart-Hub
DOCUMENTS
| General | |
|---|---|
| Product Name | PRECISE-LOG 8-Channel Current Data Logger |
| Model | PL-C |
| Inputs | |
| Connection | Pluggable terminal block for 8 external channels |
| Channels | 8 external DC current inputs |
| Programmable Range 1 | 0 - 50 mA |
| Programmable Range 2 | 4 to 20 mA |
| Resolution | 0.0018% |
| Accuracy | ±0.1% FSR @25°C |
| Load Resistor | 12 Ω |
| Over-Current Protection | Up to ±100 mA |
| Alarms | |
| Channel Alarm Thresholds | Two programmable alarm thresholds per channel |
| Alarm Indicator | Status LED flashes red during sampling when an alarm condition is active |
| On-board Memory | |
| Flash Memory | 8 MB (4MB readings) non-volatile flash memory |
| Data Retention | More than 20 years. |
| Memory-Full Options | Stop logging or FIFO mode |
| Sampling and Logging | |
| Sampling Interval | Programmable from 1 second to 12 hours |
| Logging Modes | Stop when memory is full or continue in FIFO mode |
| Logging Activation | Immediate start, delayed start, or field push-button activation |
| Communications | |
| USB Interface | Mini-USB-B |
| Included USB Cable | USB-A to Mini-USB-B |
| Communication Rate | 115200 bps |
| Battery and Power | |
| Internal Battery | Built-in 3.6 V lithium battery |
| Battery Life | Up to approximately 10 years at a 1-minute sampling interval in standalone mode |
| Software | |
| Software | SiteView (Sold separately) |
| Functions | Configuration, downloading, plotting, real-time monitoring, custom calibration and custom equations |
| Local Communication | USB |
| Computer Requirement | 1.0 GHz or faster processor; 1 GB RAM or higher; 10 GB available storage or higher |
| Required Interface | At least one available USB port |
| Physical Specifications | |
| Enclosure Material | Aluminum |
| Dimension | 88 X 64.2 X 24 mm (3.46 X 2.53 X 0.95 inches) |
| Weight | 200g |
| Mounting | Integrated holes for wall or equipment mounting |
| LED Indicators | |
| The tri-color status LED can be disabled to reduce power consumption. | |
| Green Indication | Normal sampling |
| Red Indication | Alarm condition |
| Amber Indication | Low battery |
| Operating Environment | |
| Operating Temperature | -40 ~ +70 °C (-40 °F ~ 158 °F) |
| Operating Humidity | 0 ~ 95% RH, non-condensing |
| Clock Accuracy | ±1 mintue per month |
| Approval | CE, FCC |
Logger Memory Capacity
| Sampling Interval | Enabled Channel | Logging Capacity |
|---|---|---|
| 1 minute | 1 | 8 years |
| 1 minute | 2 | 4 years |
| 1 minute | 8 | 1 year |
| 10 seconds | 1 | 485 days |
| 10 seconds | 2 | 242 days |
| 10 seconds | 8 | 60 days |
| 1 second | 1 | 48 days |
| 1 second | 2 | 24 days |
| 1 second | 8 | 6 days |
Resources: