iBG iLog Strain Gauge/Bridge Data Logger

iBG iLog Strain Gauge/Bridge Data Logger

The iBG iLOG is a compact, battery-powered strain gauge and bridge data logger designed for measuring low-level bridge signals in industrial, laboratory, research, testing, and sensor-monitoring applications.

It supports full-bridge strain gauge, load cell, pressure sensor, and other bridge-type transducer measurements. The logger provides excitation voltage, excitation-sense compensation, and a dedicated bridge input channel to help improve measurement accuracy in demanding applications.

Three models are available to support different input ranges:

  • iBG-8: ±8 mV
  • iBG-16: ±16 mV
  • iBG-32: ±32 mV

Measurement data is stored in 8 MB of non-volatile flash memory for later retrieval. The aluminum enclosure and compact iLOG design make the unit suitable for both laboratory and field applications. 

Strain Gauge and Bridge Measurement

The iBG supports 6-wire full-bridge configurations and provides:

  • 2.5 VDC excitation output
  • Excitation-sense input
  • Bridge signal input
  • Excitation-wire compensation

This makes the iBG suitable for applications using:

  • Strain gauges
  • Load cells
  • Pressure sensors
  • Force sensors
  • Torque sensors
  • Other compatible bridge-type transducers

The excitation-sense channel helps compensate for voltage drop in the excitation wiring, improving measurement accuracy when the sensor is located away from the logger.

High-Resolution Data Logging

The iBG uses a 16-bit analog-to-digital converter for high-resolution bridge measurements.

Sampling intervals are programmable from:

30 milliseconds to 12 hours

This allows the logger to support both relatively fast measurement applications and long-term standalone monitoring.

Standalone Operation and Long-Term Recording

Recorded data is stored in 8 MB of onboard non-volatile flash memory.

The logger supports:

  • Stop logging when memory is full
  • FIFO logging mode
  • Immediate logging
  • Delayed start
  • Field push-button activation

At a 1-minute sampling interval in standalone mode, the internal lithium battery provides up to 12 years of operation, depending on usage and operating conditions.

USB Communication and SiteView Software

The iBG connects directly to a Windows computer through USB.

SiteView software provides the tools required to:

  • Configure logging sessions
  • Download recorded data
  • View real-time readings
  • Display data in graph and table formats
  • Configure alarms
  • Apply custom calibration
  • Apply custom equations
  • Convert bridge measurements into engineering units
  • Export and analyze recorded data

SiteView is sold separately. 

Alarm and Excitation Control

The iBG provides configurable alarm and control functions for applications requiring more than basic data recording.

Each channel supports two programmable alarm thresholds.

The available output terminals can be configured for:

  • Alarm output
  • Excitation control

The onboard status LED provides visual indication of normal sampling, alarm conditions, and low battery status.

Remote Access and Network Integration

Although the iBG is a USB-based data logger, it can be accessed remotely over an Ethernet network by connecting it to a Microedge Smart-Hub.

Smart-Hub allows the iBG to be accessed through:

  • SiteView
  • SiteView-Web
  • SiteView SDK
  • Pluto

Smart-Hub also provides:

  • RESTful Web API
  • Modbus TCP Gateway

This allows iBG measurements to be integrated with third-party software and industrial monitoring systems.

Rugged Industrial Design

The iBG is housed in a durable aluminum enclosure and is designed for operation from:

−40°C to +70°C (−40°F to +158°F)

with up to:

95% RH, non-condensing.

Key Features

  • Full-bridge strain gauge / bridge measurement
  • 6-wire bridge support
  • Excitation voltage output
  • Excitation-sense compensation
  • Models for ±8 mV, ±16 mV, and ±32 mV input ranges
  • 16-bit analog-to-digital converter
  • 8 MB non-volatile flash memory
  • 30 ms to 12 hr programmable sampling interval
  • USB communication
  • AUX TTL-level serial communication
  • Programmable alarm thresholds
  • Stop or FIFO logging mode
  • Immediate, delayed, or push-button logging activation
  • Up to 12 years of battery life at a 1-minute sampling interval
  • Rugged aluminum enclosure
  • SiteView software support
  • Remote Ethernet access through Smart-Hub
  • RESTful Web API and Modbus TCP integration through Smart-Hub

Typical Applications

The iBG is suitable for:

  • Strain gauge measurement
  • Load cell data logging
  • Pressure sensor monitoring
  • Force measurement
  • Torque measurement
  • Bridge sensor testing
  • Structural testing
  • Laboratory research
  • Product validation
  • Long-term bridge-signal monitoring

Documents



General
Product Name iLOG Strain Gauge/Bridge data logger
Model iBG-8, iBG-16, iBG-32
Inputs
Input Connections Pluggaable terminal block for:
Excitation voltage output
Excitation sense input
Strain/Bridge input
Alarm and excitation control outputs
Bridge Configuration Support 6-wire full-bridge measurement
Excitation Output 2.5 VDC, maximum load current 5 mA
Excitation Sense Input 2.5 VDC
Bridge Input 1 bridge signal input channel
iBG-8 Input Range ±8 mV
iBG-16 Input Range ±16 mV
iBG-32 Input Range ±32 mV
Resolution 0.0018%
iBG-8 Accuracy ±0.2% FSR @ 25°C
iBG-16 Accuracy ±0.1% FSR @ 25°C
iBG-16 Accuracy ±0.05% FSR @ 25°C
Input Impedance < 1 Ω
Over-Current Protection Up to ± 100 mA
Overvoltage Protection\ Up to -3VDC ~ +40VDC
Alarm Functions
Channel Alarm Thresholds Two programmable alarm thresholds per channel
Alarm Outputs ALARM1 and A2/E terminals can be configured as alarm outputs
Alarm Output Type N-channel MOSFET switch
Alarm ON MOSFET switch ON
Alarm OFF MOSFET switch OFF
Maximum Alarm Output 200mA @ 24 VDC
Alarm Delay Programmable from 0 to 10 minutes in 1-minute increments
Alarm Indicator Status LED flashes red during sampling when an alarm condition is active
On-board Memory
Flash Memory 8 MB (4MB readings) non-volatile flash memory
Data Retention More than 20 years.
Memory-Full Options Stop logging or FIFO mode
Sampling and Logging
Sampling Interval Programmable from 20 milliseconds to 12 hours
Maximum number of enabled channels:
20 ms sampling interval: 1 channel
30 ms sampling interval: 2 channels
An external power supply is required when the sampling interval is less than 1 second.
Fast Sampling Power Requirement External power is required when the sampling interval is less than 1 second
Logging Modes Stop when memory is full or continue in FIFO mode
Logging Activation Immediate start, delayed start, or field push-button activation
Communications
USB Interface Mini-USB-B
USB Cable USB-A to Mini-USB-B cable included
AUX Interface RJ11 TTL level serial communication
Baud Rate Automatically detected from 2400 to 115200 bps on USB and AUX interfaces
Remote Network Access Available through Microedge Smart-Hub
Battery and Power
Internal Battery Built-in 3.6V Lithium Battery.
Battery Life Up to 12 years at a 1-minute sampling interval, depending on usage and operating conditions
External Power 5 VDC through USB
Software
Software SiteView (Sold separately)
Functions Logger configuration, data download, graphing, real-time monitoring, custom calibration, custom equations, alarm configuration, and data analysis
Operating Systems Windows 8, 10 and 11
Computer Requirement 1.0 GHz or faster processor; 1 GB RAM or higher; 10 GB available storage or higher
Required Interface At least one available USB port
Physical Specifications
Enclosure Material Aluminum
PCB Protection Conformal coating.
Dimension 88 X 64.2 X 24 mm (3.46 X 2.53 X 0.95 inches)
Weight 200g
Mounting Integrated holes for wall or equipment mounting
LED Indicators
The tri-color status LED can be disabled to reduce power consumption.
Green Indication Normal sampling
Red Indication Alarm condition
Amber Indication Low battery
Excitation Control
The A2/E terminal can be configured as an excitation-control output for controlling power to an externally powered sensor or device. The excitation warm-up delay is programmable from: 10 to 240 seconds in 10-second increments
Operating Environment
Operating Temperature -40 ~ +70 °C (-40 °F ~ 158 °F)
Operating Humidity 0 ~ 95% RH, non-condensing
Clock Accuracy ±1 mintue per month
Approval CE, FCC

Logger Memory Capacity

Sampling Interval Logging Capacity
1 minute 4 years
10 seconds 242 days
1 second 24 days

SiteView accesses data loggers

SiteView-Web accesses data loggers

Pluto WebServer accesses data loggers

  • $538.00(USD)

Available Options

Tags: strain gauge data logger, bridge data logger, full bridge data logger, load cell data logger, load cell recorder, pressure sensor data logger, force data logger, torque data logger, strain measurement, bridge sensor logger, USB data logger, iBG-8, iBG-16, iBG-32, iBG, iLOG