LRTD SITE-LOG Resistance Temperature Detector Data Logger

LRTD SITE-LOG Resistance Temperature Detector Data Logger

The LRTD SITE-LOG is a rugged 4-channel RTD temperature data logger designed for multi-point temperature monitoring in industrial, laboratory, research, testing, and environmental applications.

It provides four external RTD input channels plus one built-in temperature channel for ambient temperature monitoring. Three models are available for PT100, PT500, and PT1000 RTD sensors:

  • LRTD-1 – PT100
  • LRTD-2 – PT500
  • LRTD-3 – PT1000

The LRTD is a battery-powered, standalone logger with 8 MB of non-volatile flash memory, allowing temperature data to be recorded in the field and downloaded later through USB. Its rugged aluminum enclosure and conformal-coated PCB provide additional protection for long-term use in demanding environments. 

Multi-Channel RTD Temperature Measurement

The LRTD provides four external channels for resistance temperature detector measurements.

The supported resistance and temperature ranges are:

  • PT100: 20 to 400 Ω, corresponding to approximately −200°C to +850°C
  • PT500: 90 to 1600 Ω, corresponding to approximately −200°C to +630°C
  • PT1000: 185 to 3200 Ω, corresponding to approximately −200°C to +630°C

Built-in equations support RTD sensors with temperature coefficients of resistance (TCR) of:

  • 0.00385 Ω/Ω/°C
  • 0.00392 Ω/Ω/°C

Custom equations can also be used for other compatible RTD sensors.

High-Resolution Data Logging

The LRTD uses a 16-bit analog-to-digital converter for high-resolution RTD measurements.

Typical temperature resolution is:

  • PT100: 0.02°C
  • PT500: 0.015°C
  • PT1000: 0.015°C

Sampling intervals are programmable from:

20 milliseconds to 12 hours

This allows the same logger to be used for both fast testing and long-term temperature monitoring.

External power is required when the sampling interval is less than 1 second. 

Standalone Operation and Long-Term Recording

Recorded data is stored in 8 MB of onboard non-volatile flash memory and retained even when external power is removed.

Depending on the sampling interval and number of enabled channels, the logger can record from hours to years of temperature data.

The logger supports:

  • Stop logging when memory is full
  • FIFO logging mode
  • Immediate logging
  • Delayed start
  • Field push-button activation

With a 1-minute sampling interval, the internal battery can provide up to 10 years of operation, depending on usage and operating conditions.

USB Communication and SiteView Software

The LRTD connects directly to a computer through USB.

SiteView software provides the tools required to:

  • Configure logging sessions
  • Select RTD equations
  • Download recorded data
  • View real-time readings
  • Display data in graph and table formats
  • Configure alarms
  • Apply custom calibration and custom equations
  • Export and analyze recorded data

SiteView is sold separately. 

Alarm and Excitation Control

The LRTD provides configurable alarm and control functions for applications requiring more than basic data recording.

Each channel supports two programmable alarm thresholds.

The available output terminals can be configured for:

  • Alarm output
  • Excitation control

The onboard status LED provides visual indication of normal sampling, alarm conditions, and low battery status. 

Remote Access and Network Integration

Although the LRTD is a USB-based data logger, it can also be accessed remotely over an Ethernet network by connecting it to a Microedge Smart-Hub.

Smart-Hub allows the LRTD to be accessed through:

  • SiteView
  • SiteView-Web
  • SiteView SDK
  • Pluto

Smart-Hub also provides:

  • RESTful Web API
  • Modbus TCP Gateway

This allows LRTD temperature data to be integrated with third-party software, PLCs, HMIs, SCADA systems, and other industrial monitoring platforms.

Smart-Hub Specifications Page

Rugged Industrial Design

The LRTD is housed in a durable aluminum enclosure.

Its circuit board is protected with conformal coating to help resist:

  • Moisture
  • Corrosion
  • Mildew

The logger is designed for operation from:

−40°C to +70°C (−40°F to +158°F)

with up to:

95% RH, non-condensing. 

Key Features

  • 4 external RTD input channels
  • 1 built-in temperature channel
  • Available for PT100, PT500, and PT1000 sensors
  • Built-in equations for TCR 0.00385 and 0.00392
  • Custom equation support for other compatible RTD sensors
  • 16-bit analog-to-digital converter
  • 8 MB non-volatile flash memory
  • 20 ms to 12 hr programmable sampling interval
  • USB communication
  • AUX TTL-level serial communication
  • Programmable alarm thresholds
  • Alarm output and excitation control
  • Stop or FIFO logging mode
  • Immediate, delayed, or push-button logging activation
  • Up to 10 years of battery life at a 1-minute sampling interval
  • Rugged aluminum enclosure
  • Conformal-coated PCB
  • SiteView software support
  • Remote Ethernet access through Smart-Hub
  • RESTful Web API and Modbus TCP integration through Smart-Hub

Typical Applications

The LRTD is suitable for:

  • Multi-point RTD temperature monitoring
  • PT100 temperature logging
  • PT500 temperature logging
  • PT1000 temperature logging
  • Laboratory testing
  • Research and development
  • Industrial process monitoring
  • Environmental monitoring
  • Equipment testing
  • Product validation
  • Long-term temperature recording

Documents



General
Product Name SITE-LOG Resistance Temperature Detector Data Logger
Model LRTD-1 (for PT100)
LRTD-2 (for PT500)
LRTD-3 (for PT1000)
Inputs
External Input Channels 4 RTD input channels
Built-in Temperature Channel 1 onboard thermistor temperature channel
Built-in Temperature Range −40°C to +70°C (−40°F to +158°F)
Input Connection Pluggable terminal block for four external channels, excitation control, and alarm outputs
Supported RTD Models PT100, PT500, PT1000
Built-in RTD Equations TCR 0.00385 and TCR 0.00392
Input Resolution 0.0018%
Thermistor Accuracy ±0.2°C from 0°C ~ 70°C, 32°F ~ 158°F
RTD Channel Accuracy ±0.015%
LRTD-1 Accuracy / Resolution PT100: ±0.16°C @ 25°C / 0.02°C
LRTD-2 Accuracy / Resolution PT500: ±0.13°C @ 25°C / 0.015°C
LRTD-3 Accuracy Resolution PT1000: ±0.13°C @ 25°C / 0.015°C
LRTD-1 Range 20 ~ 400 Ω (PT100: -200 °C ~ 850°C)
LRTD-2 Range 90 ~ 1600 Ω (PT500: -200 °C ~ 630°C)
LRTD-3 Range 185 ~ 3200 Ω (PT500: -200 °C ~ 630°C)
Alarm Functions
Channel Alarm Thresholds Two programmable alarm thresholds per channel
Alarm Outputs ALARM1 and A2/E terminals can be configured as alarm outputs
Alarm Output Type N-channel MOSFET switch
Alarm ON MOSFET switch ON
Alarm OFF MOSFET switch OFF
Maximum Alarm Output 200mA @ 24 VDC
Alarm Delay Programmable from 0 to 10 minutes in 1-minute increments
Alarm Indicator Status LED flashes red during sampling when an alarm condition is active
On-board Memory
Flash Memory 8 MB (4MB readings) non-volatile flash memory
Data Retention More than 20 years.
Memory-Full Options Stop logging or FIFO mode
Sampling and Logging
Sampling Interval Programmable from 20 milliseconds to 12 hours
Maximum number of enabled channels:
20 ms sampling interval: 1 channel
30 ms sampling interval: 2 channels
40 ms or longer sampling interval: up to 8 channels
An external power supply is required when the sampling interval is less than 1 second.
Fast Sampling Power Requirement External power is required when the sampling interval is less than 1 second
Logging Modes Stop when memory is full or continue in FIFO mode
Logging Activation Immediate start, delayed start, or field push-button activation
Communications
USB Interface Mini-USB-B
USB Cable USB-A to Mini-USB-B cable included
AUX Interface RJ11 TTL level serial communication
Baud Rate Automatically detected from 2400 to 115200 bps on USB and AUX interfaces
Remote Network Access Available through Microedge Smart-Hub
Battery and Power
Internal Battery Built-in 3.6V Lithium Battery.
Battery Life Up to 10 years at a 1-minute sampling interval, depending on usage and operating conditions
External Power 5 VDC through USB
Software
Software SiteView (Sold separately)
Functions Logger configuration, data download, graphing, real-time monitoring, custom calibration, custom equations, alarm configuration, and data analysis
Operating Systems Windows 8, 10 and 11
Computer Requirement 1.0 GHz or faster processor; 1 GB RAM or higher; 10 GB available storage or higher
Required Interface At least one available USB port
Physical Specifications
Enclosure Material Aluminum
PCB Protection Conformal coating.
Dimension 88 X 64.2 X 24 mm (3.46 X 2.53 X 0.95 inches)
Weight 200g
Mounting Integrated holes for wall or equipment mounting
LED Indicators
The tri-color status LED can be disabled to reduce power consumption.
Green Indication Normal sampling
Red Indication Alarm condition
Amber Indication Low battery
Excitation Control
The A2/E terminal can be configured as an excitation-control output for controlling power to an externally powered sensor or device. The excitation warm-up delay is programmable from: 10 to 240 seconds in 10-second increments
Operating Environment
Operating Temperature -40 ~ +70 °C (-40 °F ~ 158 °F)
Operating Humidity 0 ~ 95% RH, non-condensing
Clock Accuracy ±1 mintue per month
Approval CE, FCC

Logger Memory Capacity

Sampling Interval Enabled Channel Logging Capacity
1 minute 1 8 years
1 minute 2 4 years
1 minute 4 2 year
10 seconds 1 485 days
10 seconds 2 242 days
10 seconds 4 141 days
1 second 1 48 days
1 second 2 24 days
1 second 4 12 days
100 milliseconds 1 4 days 20 hours
100 milliseconds 2 2 days 10 hours
100 milliseconds 4 1 day 5 hours

  • $488.00(USD)

Available Options

Tags: LRTD, RTD data logger