LPCB-1 SITE-LOG High Accuracy Current and Temperature Data Logger

LPCB-1 SITE-LOG High Accuracy Current and Temperature Data Logger

The LPCB-1 SITE-LOG is a rugged 7-channel high-accuracy DC current data logger designed for multi-point current measurement and monitoring in industrial, laboratory, research, testing, and environmental applications.

It provides seven external DC current input channels plus one built-in temperature channel for ambient temperature monitoring and temperature compensation. Each current channel can be independently configured for the required input range.

The LPCB-1 is a battery-powered, standalone logger with 8 MB of non-volatile flash memory, allowing measurement data to be recorded in the field and downloaded later through USB. Its rugged aluminum enclosure and conformal-coated PCB provide additional protection for long-term use in demanding environments. 

Multi-Channel Current Measurement

The LPCB-1 provides seven external channels for measuring DC current signals from:

  • Sensors
  • Transducers
  • Transmitters
  • Process instrumentation
  • Laboratory equipment
  • Other compatible current sources

Each channel supports programmable input ranges of:

  • 4–20 mA
  • 0–50 mA

This makes the LPCB-1 suitable for both standard 4–20 mA process signals and higher-current DC measurement applications. 

High-Resolution Data Logging

The LPCB-1 uses a 16-bit analog-to-digital converter for high-resolution current measurements.

Sampling intervals are programmable from:

20 milliseconds to 12 hours

This allows the same logger to be used for both fast testing and long-term monitoring.

According to the current LPCB-1 specification, an external power supply is required when the sampling interval is less than 1 second.

Standalone Operation and Long-Term Recording

Recorded data is stored in 8 MB of onboard non-volatile flash memory and retained even when external power is removed.

Depending on the sampling interval and number of enabled channels, the logger can record from hours to years of data.

The logger supports:

  • Stop logging when memory is full
  • FIFO logging mode
  • Immediate logging
  • Delayed start
  • Field push-button activation

With a 1-minute sampling interval, the internal battery can provide up to 10 years of operation, depending on usage and operating conditions. 

USB Communication and SiteView Software

The LPCB-1 connects directly to a computer through USB.

SiteView software provides the tools required to:

  • Configure logging sessions
  • Select current input ranges
  • Download recorded data
  • View real-time readings
  • Display data in graph and table formats
  • Configure alarms
  • Apply custom calibration and custom equations
  • Convert current signals into engineering units
  • Export and analyze recorded data

SiteView is sold separately. 

Alarm and Excitation Control

The LPTM-1 provides configurable alarm and control functions for applications that require more than simple data recording.

Each channel supports two programmable alarm thresholds.

The available output terminals can be configured for:

  • Alarm output

  • Excitation control

The onboard status LED provides visual indication of normal sampling, alarm conditions, and low battery status.

Remote Access and Network Integration

Although the LPCB-1 is a USB-based data logger, it can also be accessed remotely over an Ethernet network by connecting it to a Microedge Smart-Hub.

Smart-Hub allows the LPCB-1 to be accessed through:

  • SiteView
  • SiteView-Web
  • SiteView SDK
  • Pluto

Smart-Hub also provides:

  • RESTful Web API
  • Modbus TCP Gateway

This allows LPCB-1 current readings to be integrated with third-party software, PLCs, HMIs, SCADA systems, and other industrial monitoring platforms. Smart-Hub now replaces the older DSS-2 remote-access approach referenced on the current LPCB-1 page. 

Smart-Hub Specifications Page

Rugged Industrial Design

The LPCB-1 is housed in a durable aluminum enclosure.

Its circuit board is protected with conformal coating to help resist:

  • Moisture
  • Corrosion
  • Mildew

The logger is designed for operation from:

−40°C to +70°C (−40°F to +158°F)

with up to:

95% RH, non-condensing. 

Key Features

  • 7 external DC current input channels
  • 1 built-in temperature channel
  • Programmable 4–20 mA and 0–50 mA input ranges
  • 16-bit analog-to-digital converter
  • 8 MB non-volatile flash memory
  • 20 ms to 12 hr programmable sampling interval
  • USB communication
  • AUX TTL-level serial communication
  • Two programmable alarm thresholds per channel
  • Stop or FIFO logging mode
  • Immediate, delayed, or push-button logging activation
  • Up to 10 years of battery life at a 1-minute sampling interval
  • Rugged aluminum enclosure
  • Conformal-coated PCB
  • SiteView software support
  • Remote Ethernet access through Smart-Hub
  • RESTful Web API and Modbus TCP integration through Smart-Hub

Typical Applications

The LPCB-1 is suitable for:

  • 4–20 mA process signal logging
  • Industrial current monitoring
  • Sensor and transmitter output logging
  • Laboratory testing
  • Research and development
  • Equipment and process monitoring
  • Environmental monitoring
  • Product testing and validation
  • Long-term current recording
  • Multi-point DC current measurement

Documents



General
Product Name SITE-LOG High Accuracy Current Data Logger
Model LPCB-1
Inputs
External Input Channels 7 DC current input channels
Built-in Temperature Channel 1 onboard thermistor temperature channel
Built-in Temperature Range −40°C to +70°C (−40°F to +158°F)
Programmable Input Ranges 4–20 mA; 0–50 mA
Resolution 0.0018%
Temperature Accuracy ±0.2°C from 0°C to 70°C
Current Accuracy ±0.1% FSR @ 25°C
Load Resistor 12 Ω
Alarm Functions
Channel Alarm Thresholds Two programmable alarm thresholds per channel
Alarm Outputs ALARM1 and A2/E terminals can be configured as alarm outputs
Alarm Output Type N-channel MOSFET switch
Alarm ON MOSFET switch ON
Alarm OFF MOSFET switch OFF
Maximum Alarm Output 200mA @ 24 VDC
Alarm Delay Programmable from 0 to 10 minutes in 1-minute increments
Alarm Indicator Status LED flashes red during sampling when an alarm condition is active
On-board Memory
Flash Memory 8 MB (4MB readings) non-volatile flash memory
Data Retention More than 20 years.
Memory-Full Options Stop logging or FIFO mode
Sampling and Logging
Sampling Interval Programmable from 20 milliseconds to 12 hours
Maximum number of enabled channels:
20 ms sampling interval: 1 channel
30 ms sampling interval: 2 channels
40 ms or longer sampling interval: up to 8 channels
An external power supply is required when the sampling interval is less than 1 second.
Fast Sampling Power Requirement External power is required when the sampling interval is less than 1 second
Logging Modes Stop when memory is full or continue in FIFO mode
Logging Activation Immediate start, delayed start, or field push-button activation
Communications
USB Interface Mini-USB-B
USB Cable USB-A to Mini-USB-B cable included
AUX Interface RJ11 TTL level serial communication
Baud Rate Automatically detected from 2400 to 115200 bps on USB and AUX interfaces
Remote Network Access Available through Microedge Smart-Hub
Battery and Power
Internal Battery Built-in 3.6V Lithium Battery.
Battery Life Up to 10 years at a 1-minute sampling interval, depending on usage and operating conditions
External Power 5 VDC through USB
Software
Software SiteView (Sold separately)
Functions Logger configuration, data download, graphing, real-time monitoring, custom calibration, custom equations, alarm configuration, and data analysis
Operating Systems Windows 8, 10 and 11
Computer Requirement 1.0 GHz or faster processor; 1 GB RAM or higher; 10 GB available storage or higher
Required Interface At least one available USB port
Physical Specifications
Enclosure Material Aluminum
PCB Protection Conformal coating.
Dimension 88 X 64.2 X 24 mm (3.46 X 2.53 X 0.95 inches)
Weight 200g
Mounting Integrated holes for wall or equipment mounting
LED Indicators
The tri-color status LED can be disabled to reduce power consumption.
Green Indication Normal sampling
Red Indication Alarm condition
Amber Indication Low battery
Excitation Control
The A2/E terminal can be configured as an excitation-control output for controlling power to an externally powered sensor or device. The excitation warm-up delay is programmable from: 10 to 240 seconds in 10-second increments
Operating Environment
Operating Temperature -40 ~ +70 °C (-40 °F ~ 158 °F)
Operating Humidity 0 ~ 95% RH, non-condensing
Clock Accuracy ±1 mintue per month
Approval CE, FCC

Logger Memory Capacity

Sampling Interval Enabled Channel Logging Capacity
1 minute 1 4 years
1 minute 2 2 years
1 minute 8 181 days
10 seconds 1 242 days
10 seconds 2 121 days
10 seconds 8 30 days
1 second 1 24 days
1 second 2 12 days
1 second 8 3 days
100 milliseconds 1 58 hours
100 milliseconds 2 29 hours
100 milliseconds 8 7.2 hours

SiteView accesses data loggers

SiteView-Web accesses data loggers

Pluto WebServer accesses data loggers

  • $538.00(USD)

Tags: current data logger, DC current data logger, 7 channel current logger, multi-channel current logger, 4-20mA data logger, 4-20mA recorder, 0-50mA data logger, high accuracy current logger, process current logger, sensor data logger, transmitter data logger, USB data logger, industrial data logger, LPCB-1, SITE-LOG