LRHT-1 SITE-LOG Relative Humidity and Temperature Data Logger
The SITE-LOG LRHT-1 is a compact 2-channel, battery-powered, standalone data logger designed for accurate monitoring of relative humidity and temperature.
The logger uses an internal digital humidity and temperature sensor and provides:
- 1 relative humidity channel
- 1 temperature channel
- 8 MB non-volatile flash memory
- USB and TTL-level serial communication
- Configurable alarm and excitation-control outputs
- Long-term standalone data logging
The LRHT-1 is well suited to environmental monitoring, laboratory testing, storage monitoring, HVAC applications, research, and other applications where temperature and humidity must be recorded together.
Its aluminum enclosure and conformally coated PCB provide added protection for industrial and field use.
Relative Humidity and Temperature Measurement
The LRHT-1 simultaneously measures:
- Relative Humidity: 0–100% RH
- Temperature: −40°C to +70°C (−40°F to +158°F)
Both channels provide 12-bit measurement resolution.
Published accuracy is:
- Temperature: ±0.4°C at 25°C
- Relative Humidity: ±3% RH from 20% to 80% RH
The internal sensor design makes the LRHT-1 convenient for applications where ambient temperature and humidity are measured directly at the logger location.
Digital Humidity and Temperature Sensor
The LRHT-1 uses a digital relative-humidity and temperature sensor for environmental measurement.
Because the sensing element is integrated into the logger, no external RH probe is required.
This makes the LRHT-1 particularly suitable for:
- Room monitoring
- Storage areas
- Laboratories
- Environmental chambers
- HVAC environments
- General ambient humidity and temperature monitoring
For applications requiring the sensor to be positioned away from the logger enclosure, the LRHT-3 provides an external 3-foot temperature and humidity probe.
Standalone Operation and Long-Term Recording
The LRHT-1 operates independently after configuration and stores measurements in 8 MB of non-volatile flash memory.
The memory provides storage for approximately 4 million readings and retains recorded data for more than 20 years.
Logging can be configured for:
- Immediate start
- Delayed start
- Field push-button activation
When memory becomes full, the logger can either:
- Stop recording, or
- Continue recording in FIFO mode
The built-in 3.6 V lithium battery provides up to approximately 5 years of operation at a 1-minute sampling interval, depending on operating conditions and configuration.
USB Communication and SiteView Software
The LRHT-1 connects directly to a Windows PC through its Plug & Play USB interface.
SiteView software provides tools for:
- Logger configuration
- Data download
- Real-time viewing
- Graphing
- Alarm configuration
- Custom calibration
- Custom equations
- Data analysis
The logger also includes an AUX RJ11 interface for TTL-level serial communication.
USB and AUX communication support automatic baud-rate detection from:
2400 to 115200 bps.
Alarm and Excitation Control
The LRHT-1 provides configurable alarm and control functions for applications requiring more than basic data recording.
Each channel supports two programmable alarm thresholds.
The available output terminals can be configured for:
- Alarm output
- Excitation control
The onboard status LED provides visual indication of normal sampling, alarm conditions, and low battery status.
Remote Access and Network Integration
Although the LRHT-1 is a USB-based data logger, it can also be accessed remotely over an Ethernet network by connecting it to a Microedge Smart-Hub.
Smart-Hub acts as a communication gateway and allows connected USB data loggers to be accessed through:
- SiteView
- SiteView-Web
- SiteView SDK
- Pluto
Smart-Hub also provides:
- RESTful Web API
- Modbus TCP Gateway
This allows LRHT-1 temperature and relative-humidity data to be integrated with third-party software, PLCs, HMIs, SCADA systems, and other industrial monitoring platforms.
Rugged Industrial Design
The LRHT-1 is housed in a durable aluminum enclosure.
The PCB is conformally coated to provide additional protection against moisture, corrosion, and mildew.
Its operating environment is:
−40°C to +70°C (−40°F to +158°F)
and:
0–95% RH, non-condensing.
KEY FEATURES
Relative Humidity and Temperature Measurement
Simultaneously records RH and temperature.Internal Digital Sensor
Integrated sensor eliminates the need for an external probe.0–100% RH Measurement Range
−40°C to +70°C Temperature Range
12-Bit Resolution
Provides 12-bit measurement resolution for both RH and temperature channels.Temperature Accuracy
±0.4°C at 25°C.Relative Humidity Accuracy
±3% RH from 20% to 80% RH.8 MB Non-Volatile Flash Memory
Stores approximately 4 million readings.1 Second to 12 Hour Sampling
External power is required when the sampling interval is less than 5 seconds.Up to 5 Years Battery Life
Based on a 1-minute sampling interval.USB Communication
TTL-Level Serial Communication
AUX RJ11 interface with automatic baud-rate detection from 2400 to 115200 bps.SiteView Software Support
Programmable Alarm Thresholds
Alarm Output and Excitation Control
Remote Network Access through Smart-Hub
RESTful Web API through Smart-Hub
Modbus TCP Gateway through Smart-Hub
Rugged Aluminum Enclosure
Conformally Coated PCB
TYPICAL APPLICATIONS
The LRHT-1 is suitable for:
- Indoor environmental monitoring
- Laboratory monitoring
- Storage-room monitoring
- Warehouse monitoring
- HVAC monitoring
- Environmental chambers
- Research
- Product testing
- Manufacturing environments
- Museum and archive monitoring
- Building environmental monitoring
- Equipment-room monitoring
- Long-term temperature and humidity logging
- Remote environmental monitoring through Smart-Hub
DOCUMENTS
| General | |
|---|---|
| Product Name | SITE-LOG Relative Humidity and Temperature Data Logger |
| Model | LRHT-1 |
| Channels | 2 |
| Sensor Configuration | Internal RH and temperature sensor |
| Inputs | |
| Relative Humidity Range | 0–100% RH |
| Temperature Range | −40°C to +70°C (−40°F to +158°F) |
| Temperature Resolution | 12-bit |
| Relative Humidity Resolution | 12-bit |
| Temperature Accuracy | ±0.4°C at 25°C |
| Relative Humidity Accuracy | ±3% RH from 20% to 80% RH |
| Alarm Functions | |
| Channel Alarm Thresholds | Two programmable alarm thresholds per channel |
| Alarm Outputs | ALARM1 and A2/E terminals can be configured as alarm outputs |
| Alarm Output Type | N-channel MOSFET switch |
| Alarm ON | MOSFET switch ON |
| Alarm OFF | MOSFET switch OFF |
| Maximum Alarm Output | 200mA @ 24 VDC |
| Alarm Delay | Programmable from 0 to 10 minutes in 1-minute increments |
| Alarm Indicator | Status LED flashes red during sampling when an alarm condition is active |
| On-board Memory | |
| Flash Memory | 8 MB (4MB readings) non-volatile flash memory |
| Data Retention | More than 20 years. |
| Memory-Full Options | Stop logging or FIFO mode |
| Sampling and Logging | |
| Sampling Interval | Programmable from 1 second to 12 hours |
| Logging Modes | Stop when memory is full or continue in FIFO mode |
| Logging Activation | Immediate start, delayed start, or field push-button activation |
| Communications | |
| USB Interface | Mini-USB-B |
| USB Cable | USB-A to Mini-USB-B cable included |
| AUX Interface | RJ11 TTL level serial communication |
| Baud Rate | Automatically detected from 2400 to 115200 bps on USB and AUX interfaces |
| Remote Network Access | Available through Microedge Smart-Hub |
| Battery and Power | |
| Internal Battery | Built-in 3.6V Lithium Battery. |
| Battery Life | Up to 5 years at a 1-minute sampling interval, depending on usage and operating conditions |
| External Power | 5 VDC through USB |
| Software | |
| Software | SiteView (Sold separately) |
| Functions | Logger configuration, data download, graphing, real-time monitoring, custom calibration, custom equations, alarm configuration, and data analysis |
| Operating Systems | Windows 8, 10 and 11 |
| Computer Requirement | 1.0 GHz or faster processor; 1 GB RAM or higher; 10 GB available storage or higher |
| Required Interface | At least one available USB port |
| Physical Specifications | |
| Enclosure Material | Aluminum |
| PCB Protection | Conformal coating. |
| Dimension | 88 X 64.2 X 24 mm (3.46 X 2.53 X 0.95 inches) |
| Weight | 200g |
| Mounting | Integrated holes for wall or equipment mounting |
| LED Indicators | |
| The tri-color status LED can be disabled to reduce power consumption. | |
| Green Indication | Normal sampling |
| Red Indication | Alarm condition |
| Amber Indication | Low battery |
| Excitation Control | |
| The A2/E terminal can be configured as an excitation-control output for controlling power to an externally powered sensor or device. The excitation warm-up delay is programmable from: 10 to 240 seconds in 10-second increments | |
| Operating Environment | |
| Operating Temperature | -40 ~ +70 °C (-40 °F ~ 158 °F) |
| Operating Humidity | 0 ~ 95% RH, non-condensing |
| Clock Accuracy | ±1 mintue per month |
| Approval | CE, FCC |
Logger Memory Capacity
| Sampling Interval | Enabled Channel | Logging Capacity |
|---|---|---|
| 1 minute | 1 | 4 years |
| 1 minute | 2 | 2 years |
| 10 seconds | 1 | 242 days |
| 10 seconds | 2 | 121 days |
| 1 second | 1 | 24 days |
| 1 second | 2 | 12 days |
Resources: